Fan Failure Detect Circuit

This circuit was designed for a Flourescine Angiography Light Box.

Flourescine Angiography is used for the detection of bleeding in the eye.

This Light Box contains a Xenon Lamp that is used to illuminate the inside

of a patients eye.

The Xenon Lamp can overheat without proper cooling.

In this case, the lamp should be turned off and allowed to cool.

One condition that can cause overheating is the failure of the cooling fan.

This circuit is designed to detect such a failure and turn off the Xenon Lamp.

This circuit uses a "Microchip TC-670" chip for failure detection.

 

TC670 Data Sheet

PADS Board Layout

   

 

This is the original breadboarded circuit
First test with lamp and power supplies
Must have a cover to protect your eyes
First test with a real lightbox
First Prototype Board from ExpressPCB
First professionally made board
Tested and working in the FA Lightbox